Description
LeCroy has a bandwidth of up to 45GHz by utilizing widely adopted and proven SiGe processes, custom 2nd generation ASIC designs, and 6th generation Digital Bandwidth Interleave (DBI) technology to achieve unprecedented real-time oscilloscope performance:
45 GHz
120 GS/s
768 Mpts/Ch Analysis Memory
20 GHz/4ch Model
Four channel performance at 20 GHz is pristine with exceptional rise time, overshoot, total and random jitter noise floor, and electrical noise performance. SDAII Serial Data Analysis Software LeCroy’s SDAII software is optimized for fast eye diagram creation on long records, and deep insight of jitter pathologies with more parameters and analysis views to help you understand the root cause of high jitter.
Additional Capability Standard
Key Features
- Up to 30 GHz bandwidth and
80 GS/s sample rate - The industry’s only true hardware
14.1 Gb/s serial trigger - Low Jitter Measurement Floor and
exceptional timebase stability - Comprehensive set of serial data
analysis, debug, validation and
compliance tools - Integrated 50 Ω and 1 MΩ inputs
for true connection and probing
flexibility - Integrated standard and custom
measurements and math functions
for unrivaled analysis capability - Multi-lane serial data eye, jitter and
crosstalk analysis - Real-time de-embedding,
emulation, and equalization
Specification
Vertical System |
|
Analog Bandwidth (Max) | 13 GHz |
Analog Bandwidth @ 50 Ω (-3 dB) (ProLink Input) | 13 GHz (≥10 mV/div) |
Analog Bandwidth @ 50 Ω (-3 dB) (ProBus Input) | 3.5 GHz (≥10 mV/div) |
Analog Bandwidth @ 1 MΩ (-3 dB) (ProBus Input) | 500 MHz (typical, ≥2 mV/div) |
Rise Time (10-90%, 50 Ω) | 32.5 ps (test limit, flatness mode) |
Rise Time (20-80%, 50 Ω) | 24.5 ps (flatness mode) |
Input Channels | 4 (Any combination of ProLink and ProBus inputs) |
Bandwidth Limiters | 20 MHz, 200 MHz, 1 GHz, 4 GHz, 6 GHz, 8 GHz |
Input Impedance | ProLink Inputs: 50 Ω+/-2% for ≤100 mV/div, 50 Ω+/-3% for >100 mV/div ProBus Inputs: 50 Ω+/-2% or 1 MΩ||16pF, 10 MΩ || 11 pF with supplied Probe |
Input Coupling | ProLink Inputs – 50 Ω: DC, GND ProBus Inputs – 1 MΩ: AC, DC, GND; 50 Ω: DC, GND |
Maximum Input Voltage | 50 Ω (ProLink): ±2 Vmax@≤100mV/div, 5.5Vrms@>100mV/div 50 Ω (ProBus): ±5 Vmax, 3.5 Vrms 1 MΩ (ProBus): 250 V max. (peak AC: < 10 kHz + DC) |
Channel-Channel Isolation | DC to 10 GHz: 50 dB (>315:1) 10 to 15 GHz: 46 dB (>200:1) 15 to 20 GHz: 40 dB (>100:1) (For any two ProLink input channels, same or different v/div settings, typical) |
Vertical Resolution | 8 bits; up to 11 bits with enhanced resolution (ERES) |
Sensitivity | 50 Ω (ProLink): 2 mV-1 V/div, fully variable (2-9.9 mV/div via zoom) 50 Ω (ProBus): 2 mV-1 V/div, fully variable; 1 MΩ (ProBus): 2 mV-10 V/div, fully variable |
DC Vertical Gain Accuracy (Gain Component of DC Accuracy) | ±1% F.S. (typical), offset at 0V; ±1.5% F.S. (test limit), offset at 0V |
Vertical Noise Floor (50 mV/div) | 1.80 mVrms (typical) |
Offset Range | 50 Ω (ProLink): ±500 mV @ 2-100 mV/div ±4 V @ >100 mV/div -1 V/div 50 Ω (ProBus): ±750 mV @ 2-100 mV/div ±4 V @ >100 mV/div -1 V/div 1 MΩ: ±1V @ 2-140 mV/div ±10V @ 142mV-1.40V/div ±100V @ 1.42V-10V/div |
DC Vertical Offset Accuracy | ±(1.5% of offset setting + 1.5% F.S. + 1 mV) (test limit) |
Reviews