Description
Features and Specifications of Keysight (Agilent/HP) 8341B:
- Range: 10 MHz to 20.0 GHz
- Resolution: n x 1 Hz, Where N = harmonic multiplication (1 to 3). Refer to Frequency ranges and Bandswitch Points description below.
- Accuracy: Same as Time Base Accuracy
- Frequency: 10 MHz to 20.0 GHz
- Aging Rate: 1 X 10 9 per day, 2.5 X 10 7 per year after 72 hour warm up if HP 8341B has been disconnected from ac power for less than 24 hours. Aging rate is achieved after 7 to 30 days warm up if HP 8341B has been disconnected from ac power for greater than 24 hours.
- Temperature Coefficient: Typically < 1 X 10 10 per °C
- Change due to 10% line voltage change: Typically < 1 X 10-11
- Accuracy: Overall accuracy of internal time base is a function of time base calibration ± aging rate ± temperature effects ± line effects
- Switching Time: <50 msec
Description
Specifications are the performance standards, or limits, against which the instrument may be tested. The following Specifications apply for temperatures between 0 and +50’C except were noted. Specifications apply with the PEAK function ON in the CW and MANUAL modes of operation, and with periodic use of AUTO TRACKING CALIBRATION in swept operation. Supplemental Performance Characterisitcs are in italics in this table and are intended to provide information useful in applying the instrument by giving typical, but non-warranted, performance parameters. These are denoted as “typical,” “nominal,” or “approximate.”