TestEquity 1007H Features:

  • Temperature -73°C to +175°C
  • Humidity 10% to 95% RH.
  • 7 Cu Ft Workspace, 24 W x 21 H x 24 D (198 liters)
  • Programmable Temperature/Humidity Controller
  • RS-232 Interface (Opt. GPIB and Ethernet)
  • High/Low Limit Control and Alarm
  • Viewing Window & Interior Light
  • 4″ Access Ports on Left & Right Side
  • Non-CFC Cascade Refrigeration

Contact TestWorld to get the best pricing on a used/refurbished TestEquity 1007H Temperature/Humidity Chamber. Rental and lease options available.

Programmable Temperature/Humidity Controller
Up to 256 steps can be programmed into as many as 40 nameable profiles. A large LED readout indicates the actual chamber temperature with 0.1° resolution. The context sensitive information key and guided steps make profile programming fast and easy. A four-line backlit LCD displays relative humidity with 0.1% RH resolution, programming, setup, operating and help information. Internal logic provides refrigeration compressor control for responsive and reliable performance. Includes event outputs to control remote devices. RS-232C interface is also included. GPIB, Ethernet and analog retransmit outputs are available as options.

High/Low Temperature Limit Controller and Alarm
Provides independent protection against excess temperatures. Both high and low limits can be set. The upper LED readout indicates the actual chamber temperature while the lower readout indicates the status The limit controller will shut down the chamber and trigger the audible alarm in the event of an out of limit condition. A set of uncommitted contacts are available to provide a safety power interlock for your test sample.

Accurate Humidity Measurement
A fast responding Vaisala capacitive-type sensor eliminates the need for unreliable wet-wicks.

Viewing Window
Heated multi-pane window eliminates fog and condensation without a manual wiper.

Two 4″ Access Ports
Lets you attach wires and sensors to your test sample through both the left and right side of the chamber.

Adjustable Shelf
A heavy-duty stainless steel wire shelf ensures proper airflow around your test sample for uniform temperature distribution.

Built-In Refrigeration Gauges
Provides system status for easy preventive maintenance and service.

Casters
Rugged 3″ casters permit easy movement.

Accurate Humidity Measurement
A fast responding capacitive-type sensor eliminates the need for unreliable wet-wicks.

Reliable Vapor Generator
Unlike “atomizing” systems, the heated vapor generator has no pump to fail or orifice to clog. The stainless-steel vapor generator assembly is double-protected against low-water and over-temperature conditions. A float switch senses low-water level. The immersion heater has an embedded thermocouple for ultimate thermal protection.

No-Sweat Cabinet
We use a combination of structural foam and fiberglass blanket insulation for a higher combined R-factor than just fiberglass alone. The door sealing surface is kept warm by circulating excess heat from the refrigeration system. The result is a cabinet that does not sweat due to condensation during low temperature operation.

Heated Viewing Widow
Eliminates fog and condensation without a manual wiper.

Water Recirculation System
The optional water recirculation system provides a reliable source of demineralized water for the humidity system, using ordinary tap or pre-filtered water. Condensate from the chamber is returned back to the water tank. The recirculation pump operates only when needed to fill the chamber’s humidifier to the correct level. Perfect for installations where a water line and drain are not available.
(Optional, and highly recommended) More Info >>

Reliability & Safety That’s Designed In
Non-CFC refrigerants are used in a cascade (two compressors) configuration. Expansion valves provide optimum performance throughout the operating range. Pressure regulators keep the operating pressures within safe limits regardless of load conditions. Liquid injection ensures cool compressor operation during high-temperature pull down for long life. Sequential starting of each compressor reduces the current demand on start-up. Each compressor is protected by high/low pressure switches. Care has been taken to use a minimum number of fittings, joints and welds to reduce the possibility of refrigerant leaks.

All electrical control components are UL approved. Each branch circuit is individually protected with class-CC fuses. A main power disconnect switch is provided on the front panel. The nichrome air-heater has a low watt-density and even heat distribution for reliable performance and fast response. A fusible link provides fail-safe protection against thermal runaways, in addition to the microprocessor-based high/low limit controller.

The side and rear panels are removable to permit complete access to the refrigeration and humidity system for ease of service and preventive maintenance.

 