Description
Teledyne LeCroy – AP-1M 1 MOhm Adapter provides a high impedance (1 Mohm) interface for use on the WaveMaster family of products (DSO, DDA, SDA). The Adapter presents a high impedance path (instead of the 50 ohm WaveMaster input impedance) and allows users to measure higher voltage signals while also expanding the full-scale range capability of measurement. The specifications shown pertain only to the AP-1M Adapter itself. Users can achieve even higher full-scale ranges and offset ranges with the overall signal measurement chain (WaveMaster family product + AP-1M Adapter + Probe), depending on the probe used. The AP-1M Adapter is supplied with a PP005A passive probe, but it is suitable for use with other probe types, including current probes, which require a 1 Mohm path. The AP-1M also serves as a ProLink-to-ProBus adapter.
Teledyne LeCroy – AP-1M Specifications
- Bandwidth: 500 MHz typical, 350 MHz guaranteed
- Input Impedance: 1 Mohm
- Full-scale Range: +/-8 V pk (+/-80 V pk with PP005A probe)
- Max. Withstand Voltage: +/-150 V pk Cat I
- Input Capacitance: 20 pF +/-2 pF With PP005A: 11 pF at probe tip
- Coupling: DC, AC (cutoff < 10 Hz)
- Gain accuracy: +/-3% system accuracy with AP-1M Adapter used with
- WaveMaster family products
- Offset Accuracy: +/-(3% of full scale + 1.5% of offset value + 2 mV)
- Certification: CE Approved, Conforms to EN 61010-1
- Coupling: DC, AC (cutoff < 10 Hz)
- Gain accuracy: ±3% system accuracy with AP-1M
- Adapter used with WaveMaster family products
- Offset Accuracy: ±(3% of full scale + 1.5% of offsetvalue + 2 mV)
- Meets or exceeds Environmental Specifications
- (Operating Temperature: 5 °C to 40 °C, etc.) of
- WaveMaster family products.
- Certification: CE Approved, Conforms to EN 61010-1(Safety) and EN 61326-1 (EMC)
Reviews
There are no reviews yet.