Fluke (Hart Scientific) 9100S/9102S Handheld Dry-Well Calibrators

Fluke (Hart Scientific) 9100S/9102S Handheld Dry-Well Calibrators

Description

Overview Fluke 9100S Dry-Well:

Since Fluke (Hart Scientific) introduced the world’s first truly handheld dry-well, many have tried to duplicate it—in vain. Despite its small size (2¼ inches high and 5 inches wide) and light weight, the Fluke 9100S outperforms every dry-well in its class in the world.

It’s simple and convenient, too. Anyone can learn to use one in less than 15 minutes. It has a range to 375°C (707°F) and is perfect for checking RTDs, thermocouples, and small bimetal thermometers in the field.

Plug it in, switch it on (Note, these wells are fixed and not interchangeable,) set the temperature with the front-panel buttons, and insert your probe into the properly sized well. Compare the reading of your device to the display temperature or to an external reference, and the difference is the error in your device. With a proprietary Fluke Calibration temperature controller, the Fluke 9100S has a display resolution of 0.1 degrees, an accuracy of ±0.5°C, a stability of ±0.1°C, and a stabilization time of just five minutes.

Overview Fluke 9102S Dry-Well:

For work in the temperature range of –10°C to 122°C, the Fluke 9102S dry-well is another first in the industry, featuring stability to ±0.05°C.

This dry-well is only four inches high and six inches wide, achieves temperatures as low as –10°C, includes a NIST-traceable calibration, and is stable to ±0.05°C. The Fluke 9102S is excellent for dial gauges, digital thermometers, bulb switches, and other sensors that need calibration below ambient.

The Fluke 9102S has two wells so you can use one for a reference thermometer to increase accuracy. Both wells are 1/2 inch in diameter, and each has inserts available for almost any sensor size. The 9102S also has a battery pack option that gives you approximately four hours of field use when AC power is unavailable.

Specifications Fluke 9100S Fluke 9102S
Range 35 °C to 375 °C (95 °F to 707 °F –10 °C to 122 °C (14 °F to 252 °F) at23 °C ambient
Accuracy ± 0.25 °C at 50 °C; ± 0.25 °C at 100 °C; ± 0.5 °C at 375 °C ± 0.25 °C
Stability ± 0.07 °C at 50 °C; ± 0.1 °C at 100 °C; ± 0.3 °C at 375 °C ± 0.05 °C
Well-to-Well Uniformity ± 0.2 °C with sensors of similar size at equal depths within wells
Heating Times 35 °C to 375 °C: 9.5 minutes ambient to 100 °C: 10 minutes
Stabilization 5 minutes 7 minutes
Cooling Times 375 °C to 100 °C: 14 minutes ambient to 0 °C: 10 minutes
Well Depth 102 mm (4 in);1.6 mm (1/16 in) hole is 89 mm (3.5 in) deep 102 mm (4 in)

Contact TestWorld to get the best pricing on a used/refurbished Fluke (Hart Scientific) 9100S/9102S Handheld Dry-Well Calibrators. Rental and lease options available.