Contact us: 1-855-200-TEST (8378) or sales@testworld.com

Astro-Med Dash 18X 18 Channel Data Recorder

Astro-Med Dash 18X 18 Channel Data Recorder

Categories: ,

Description

The Astro-Med Dash 18X 18 Channel Data Recorder gives users the functionality of a scope, meter, data recorder, datalogger and audio recorder in one system with the footprint of a laptop computer with 18 universal inputs for voltage, thermocouple and bridge signals. The Astro-Med Dash 18X’s large real-time waveform display, internal hard drive for capturing data, integral PC connectivity and optional chart recorder, give users the ability to view data in any format that they want. With audio and video snapshot recording – synchronized to waveform data – the Astro-Med Dash 18X ensures that users will capture all aspects of the test. The Astro-Med Dash 18X captures 18 channels of data continuously to a dedicated hard drive at a sample rate of 100 kHz per channel, allowing users to capture the details in high-speed glitches or transients. For long-term trend recording, the Dash 18X allows users to set the sample rate as low as one sample per minute for days, weeks or even months of recording. The flexible Dash 18X allows users to choose the sample rate that is best for the application.

Three different data acquisition modes of the Astro-Med Dash 18X 18 Channel Data Recorder make it adaptable for any application. Real-time recording mode displays waveforms while capturing to the hard drive. Signals are displayed on the large, color display, giving instant data visibility. Scrolling waveforms show exactly what signals are doing in real-time. Use the scope mode for high-speed signals requiring greater timebase resolution. Scope mode gives a detailed, real-time view of high frequency signals while capturing data to the hard drive in the background. Traditional scope controls give the quickest and easiest way to look at high-speed data. Acquire/review mode is ideal for applications where immediate access to captured data is important. It allows users to look back at previously captured data without interrupting real-time recording.

Contact TestWorld today for the best price on a Astro-Med Dash 18X 18 Channel Data Recorder.

Explore Financial options to find a cost management solution best for your company:

Astro-Med Dash 18X 18 Channel Data Recorder Specifications:

Color Display
Type Active matrix color LCD (TFT) with 17″ viewing area
Resolution 1280 x 1024
Touch Full-screen, resistive
Functions User interface with touch-based icons and menus; Real-time waveform monitoring; Review previous waveform records while recording; Overlay numeric values in Engineering Units
Interface
Ethernet 10/100BaseT
DVD Supports archive of data to
CD-R, CD-RW, DVD-R, DVD+R & DVD+RW
VGA For displaying data on an external monitor
USB 1.1 For external peripherals and file export
Analog Input Boards
Maximum Boards 3
Maximum Waveforms 18
A/D 14 bit
Standard Event Inputs
Number of Inputs 8
Connector D-shell (9-pin)
Input Type TTL with pull-up (0 to 5VDC)
Response Detects if duration is greater than 0.01 msec
Real-time Signal Processing
Filters Low Pass Stops: 1 Hz to 10,000 Hz; High Pass Starts: 0.1 Hz to 100 Hz; Notch: 50 or 60 Hz center
RMS Time constant selectable from 0.02 to 2 seconds
Frequency to Voltage 5 Hz to 20 kHz, +/- 0.05% of measurement + 0.1 Hz
Cross-Channel Sum, Difference, Ratio, Product (Power)
Real-time Signal Testing
Basic All active signals monitored simultaneously
Test Types Tests include window, slope/level, slew and event pattern
Outputs Result available for alarms, trigger, abort and external TTL
Data Capture
Recording Method Internal 73 Gbyte hard drive
Max. Rate 100,000 samples/second per channel (all channels)
Min. Rate 1 sample/minute
Dual Rate Second slower rate available when no trigger is used
Total Capacity 16 billion samples
Max. Record 2 billion samples
Header Information on units, range, sample rates, etc. saved with data
Events All standard event inputs captured with waveforms
Trigger Point Lock Amounts of pre- and post-trigger is user adjustable
Auto Re-arm Allows automatic stacking of captures
Auto Playback Yes
External Sample Rate TTL sample clock up to 40,000 Hz
Signal Conditioner Specifications (General)
Channels per Module 6
Isolation 250 Vrms (iso-common to chassis and other iso-commons)
Bandwidth 12 kHz (-3dB)
Input Coupling DC
Zero Suppression Yes
User Engineering Units Yes
Calibration Semi-automated to external reference
Single Ended Voltage Input
Input Type Isolated, single-ended
Connector Guarded banana jack
Max. rated input +/- 250 Vrms
Max. transient input +/- 400 V
Specified ranges 40 to 400 VFS; 4 to 40 VFS; 0.4 to 4VFS
Accuracy (25°C) +/- 0.5% of attenuator
Min. Input Impedance 1 Megohm
Differential Voltage Measurements (use for bridge measurements)
Input Type Isolated, differential
Connector Screw terminal header
Absolute Max. Input +/- 40 V differential
Measuring Ranges 200 to 1600 mVFS; 50 to 500 mVFS; 5 to 50 mVFS
Accuracy (25°C) +/- 0.5% of attenuator
Min. Input Impedance 200 kOhm
CMR at 60 Hz > 60dB
Auto Balance Yes, limited to maximum span
Thermocouple Measurements
Input Type Isolated, differential with screw terminal header connector
Absolute Max. Input +/- 40 V
Specified Range Type J: 0 to 760 °C
Type K: 0 to 1370 °C
Type T: -160 to 400 °C
Type E: -100 to 870 °C
Type N: 0 to 1300 °C
Accuracy (25°C) 0.5% of measurement + 2 °C
Bandwidth 10 Hz (-3dB)
Intrinsic Noise 0.5 °C (1 Hz filter)
Linearization NIST polynomial
Units °C or °F
Excitation
Connector Screw terminal header
Excitation Isolated 10V @ 20 mA
Adjustable 0.1 to 10 V
Frequency Measurements
Method Zero-crossing, channels 1, 7 and 13 only
Signal Amplitude Peak-to-peak minimum is 20% of attenuator
Hysteresis Yes
Removable Drive
Type DVD+R
Format MSDOS® format
Function Setup files, software upgrades, data transfer/archive
Power
Input voltage range 100 to 250 VAC
Frequency Range 47 Hz to 63 Hz
Power Consumption 300 W maximum
Compliance
Safety EN 61010-1
EMC FCC Class A, EN 61326
Power Harmonics IEC1000-3-2
Physical
Enclosure Aluminum
Dimensions 16.1″ L x 15.1″ W x 6.4″ H
Weight 29 lbs.
Environmental
Operating Temperature 5 to 40 °C (40 to 105°F)
Non-operating Temperature -20 to 60 °C (-4 to 140 °F)
Operating humidity 10% to 90%, non condensing
Optional Chart Recorder – Model SCR-18
Chart Width 11″
Max. Chart Speed 50 mm/sec
Optional DC Input Power – Model PIM-18
Input Range 9-18 VDC
Input Power 200 W max
Fuse User replaceable 25 A
Optional DC Input Power – Model PIM-18A
Input Range 20-32 VDC
Input Power 240 W max
Fuse User replaceable 10 A

Reviews

There are no reviews yet.